Enhancement of carbon detection sensitivity in laser induced breakdown spectroscopy with low pressure ambient helium gas K.H. Kurniawan, M.O. Tjia, K. Kagawa, D.P. Kurniawan, M M Suliyanti, R. Hedwig, Z.S. Lie, E. Jobiliong, M. Pardede, N. Idris Gems Andor ICCD Czerny-Turner (Unspecified)
Laser: Nd:YAG
1064.0000nm
60.0000mJ
10.000Hz
Gate Delay: 0.200us
Gate Width: 50.000us
The authors utilise low pressure He atmosphere to enable the analysis of C in Jasper, with minimal damage to the samples. While an ANDOR Mechelle is used for the spectra work, they switch to a McPherson Czerny-Turner spectrograph for the calibration curve development. Results show that 2.6kPa He gives significantly stronger C peaks than 0.5kPa Air (8-10x). Calibration curves were prepared using Pellets of KBr and CaCO3
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
C 0.6000 (Calibration Curve Slope) 247.8000 N/A Univariate 5.0000-100.0000 Minimum concentration sample contained 5ppm C. Detection limited inferred from usual 3 sigma/slope


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes