LIBS Info: Element Analysis

Title Authors Material Detector Spectrometer Software
The influence of multivariate analysis methods and target grain size on the accuracy of remote quantitative chemical analysis of rocks using laser induced breakdown spectroscopy [PLS] Rhonda McInroy, Trevor G. Graff, Stanley A. Mertzman, Seth D. Humphries, James F. Bell III, Roger C. Wiens, Richard V. Morris, Ryan B. Anderson, S.M. Clegg Geological Ocean Optics HR2000 Ocean Optics HR2000
Laser: Nd:YAG
1064.0000nm
17.0000mJ
10.000Hz
Gate Delay: Noneus
Gate Width: Noneus
The sample data was collected on an instrument designed to simulate the ChemCam on Mars - 7m standoff LIBS, 7 Torr CO2. The researchers used Chemometrics/Multivariate Analysis [PLS] and several types of Neural Network [ANN]
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes
K 0.8700 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs. Feature Selection used to provide best performance.
Na 1.0600 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs. Feature Selection used to provide best performance.
Ca 1.6700 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs.
Mg 2.3700 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs. Feature Selection used to provide best performance.
Mn 0.4900 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs. Feature Selection used to provide best performance.
Fe 4.8200 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs. Feature Selection used to provide best performance.
Al 4.1400 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs.
Ti 0.6100 % -10.0000 PLS 7 Torr CO2 Atmosphere. Rock Slabs. Feature Selection used to provide best performance.
Si 4.8800 % -10.0000 PLS 7 Torr CO2 atmosphere, on pressed pellets and rock slabs.