LIBS Info: Element Analysis

Title Authors Material Detector Spectrometer Software
Analytical control of liquid steel in an induction melting furnace using a remote laser induced plasma spectrometer S. Palanco, S. Conesa, J.J. Laserna Steel ICCD Not Known
Laser: Nd:YAG
1064.0000nm
350.0000mJ
20.000Hz
Gate Delay: Noneus
Gate Width: Noneus
LIBS is performed on molten steel using effectively a StandOff LIBS rig. Analysis distance was ~5m Authors note that the intensity of the spectra increases with increasing steel temperature. Their investigations indicate that both plasma temperature increases and higher ablated mass contribute to this effect
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
Ni 540.0000 (Stated) 341.4760 N/A Univariate Calibration 1900.0000-146650.0000 RSD=1.83%. Cr also analysis (LOD=1190ppm) but no wavelength details supplied


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes