Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Laser induced breakdown spectrometry of vanadium in titania supported silica catalysts | P. Lucena, L.M. Cabalin, E. Pardo, F. Martin, L.J. Alemany, J.J. Laserna, M. Sabsabi | Titania | Stanford Computer Optics, 4 Quik 05 | Chromex 500 IS | |
Laser: | Nd:YAG 532.0000nm 4.0000mJ NoneHz |
||||
Gate Delay: | 1.300us | ||||
Gate Width: | 1.000us | ||||
Original materials were ground and then pressed into tables for analysis. The paper also contains a study on the effect of the focusing conditions on the Signal to Background ratio. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
V | 38.0000 (Stated) | 411.1800 | N/A | Univariate Calibration | 200.0000-10000.0000 | Calibration Curve was constructed using the V 411.18nm peak normalised to the Ti peak at 407.85nm |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|