LIBS Info: Element Analysis

Title Authors Material Detector Spectrometer Software
Laser induced breakdown spectrometry of vanadium in titania supported silica catalysts P. Lucena, L.M. Cabalin, E. Pardo, F. Martin, L.J. Alemany, J.J. Laserna, M. Sabsabi Titania Stanford Computer Optics, 4 Quik 05 Chromex 500 IS
Laser: Nd:YAG
532.0000nm
4.0000mJ
NoneHz
Gate Delay: 1.300us
Gate Width: 1.000us
Original materials were ground and then pressed into tables for analysis. The paper also contains a study on the effect of the focusing conditions on the Signal to Background ratio.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
V 38.0000 (Stated) 411.1800 N/A Univariate Calibration 200.0000-10000.0000 Calibration Curve was constructed using the V 411.18nm peak normalised to the Ti peak at 407.85nm


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes