LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Evaluating laser induced breakdown spectroscopy sensor technology for rapid source characterization of rare earth elements | Dustin L McIntyre, Jinesh C Jain, Chet R Bhatt, Daniel A. Hartzler | Liquids | Andor ICCD | Andor Shamrock 500i | |
Laser: | Nd:YAG 1064.0000nm 4.2000mJ 5.000Hz |
||||
Gate Delay: | 0.200us | ||||
Gate Width: | 3.000us | ||||
The paper uses a prototype LIBS system using a diode pumped, passively Q-switched laser for the analysis of Rare Earth Elements in both solid (pellets) and liquid samples. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
Eu | 9.9000 (Calibration Curve Slope) | 412.9400 | 466.36nm, 462.72nm 459.40nm, 420.51nm, 390.69nm, 382.97nm | Univariate | 250.0000-2500.0000 | Analysis on pellet samples. The authors note several interferences to Al and Ca, that marred the performance. Detection limits in the range 10-39ppm obtained from the other peaks. |
Yb | 1.1500 (Calibration Curve Slope) | 398.8000 | N/A | Univariate | 100.0000-1000.0000 | Analysis in Liquids. 369.42nm peak was considered, but showed strong self absorption |
Eu | 0.8500 (Calibration Curve Slope) | 459.4000 | 466.19nm, 462.72nm | Univariate | 100.0000-1000.0000 | Analysis in liquids. Detection limits of 1.05-1.54ppm obtained with the other peaks |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|