From Calibration-Free to Fundamental Parameters Analysis: A comparison of three recently proposed approaches [One Point Calibration] E. Grifoni, S. Legnaioli, Giulia Lorenzetti, S. Pagnotta, F. Poggialini, V. Palleschi Brass MODI (MObile Dual-pulse Instrument) AvaSpec-2048-SPU
Laser: Nd:YAG
1064.0000nm
10.0000mJ
NoneHz
Gate Delay: Noneus
Gate Width: Noneus
This paper compares the performance of 3 Calibration-Free/Fundamental Parameters Analysis algorithms for the analysis of 6 Brass Standards. Brass was chosen as it is known to have strong self-absorption effects in the spectra. The algorithms compared are One Point Calibration, Calibration-Free Inverse method and Csigma analysis.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes
Pb 8475.4548 ppm -10.0000 One-Point Calibration Free
Cu 31302.2896 ppm -10.0000 One-Point Calibration Free
Zn 20904.5450 ppm -10.0000 One-Point Calibration Free
Sn 18484.2275 ppm -10.0000 One-Point Calibration Free