LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Quantitative detection of metallic traces in water-based liquids by microwave-assisted laser- induced breakdown spectroscopy | Matthew Wall, Zhiwei Sun, Zeyad T. Alwahabi | Liquids | Andor iStar III | Andor Shamrock 500i | |
Laser: | Nd:YAG 532.0000nm NonemJ 10.000Hz |
||||
Gate Delay: | 0.250us | ||||
Gate Width: | 700.000us | ||||
In this paper, a horizontal liquid jet is analysed for Indium. The detection limits are significantly enhanced by the use of 1.2kW microwave source. Largest enhancement occurs at a laser fluence of 85J/cm^2, with enhancement decreasing with increasing laser fluence. Detection limits for LIBS and Microwave Assisted LIBS are presented |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
In | 10.8000 (Calibration Curve Slope) | 451.1300 | N/A | Univariate Calibration | 100.0000-1005.0000 | Microwave-assisted LIBS in solution |
In | 124.0000 (Calibration Curve Slope) | 451.1300 | N/A | Univariate Calibration | 100.0000-1005.0000 | Standard LIBS in liquids |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|