LIBS Info: Element Analysis

Title Authors Material Detector Spectrometer Software
Quantitative detection of metallic traces in water-based liquids by microwave-assisted laser- induced breakdown spectroscopy Matthew Wall, Zhiwei Sun, Zeyad T. Alwahabi Liquids Andor iStar III Andor Shamrock 500i
Laser: Nd:YAG
532.0000nm
NonemJ
10.000Hz
Gate Delay: 0.250us
Gate Width: 700.000us
In this paper, a horizontal liquid jet is analysed for Indium. The detection limits are significantly enhanced by the use of 1.2kW microwave source. Largest enhancement occurs at a laser fluence of 85J/cm^2, with enhancement decreasing with increasing laser fluence. Detection limits for LIBS and Microwave Assisted LIBS are presented
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
In 10.8000 (Calibration Curve Slope) 451.1300 N/A Univariate Calibration 100.0000-1005.0000 Microwave-assisted LIBS in solution
In 124.0000 (Calibration Curve Slope) 451.1300 N/A Univariate Calibration 100.0000-1005.0000 Standard LIBS in liquids


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes