LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Detection of trace concentrations of helium and argon in gas mixtures laser-induced breakdown spectroscopy | E.D. McNaghten, A.M. Parkes, B.C. Griffiths, A.I. Whitehouse, S. Palanco | Gases | Oriel InstaSpec V | Oriel MS125 | |
Laser: | Nd:YAG 1064.0000nm 850.0000mJ 10.000Hz |
||||
Gate Delay: | Noneus | ||||
Gate Width: | Noneus | ||||
Rare paper measuring noble gas concentrations in other gas mixtures (rather than using the He or Ar as a buffer gas for other analysis). Detection limits are strongly dependent on the gas mix. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
Ar | 4.9000 (Calibration Curve Slope) | 763.5100 | N/A | Univariate | -10.0000--10.0000 | Detection limit for Argon in Helium. LODs in the range 4.6-270ppm depending on main diluent gas (Nitrogen, Helium) |
He | 46.0000 (Calibration Curve Slope) | 587.5600 | N/A | Univariate | -10.0000--10.0000 | Helium in Nitrogen. LODs in the range 46-160ppm depending on diluent gas (Nitrogen or Argon) |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|