Fast and sensitive trace metal analysis in aqueous solutions by laser-induced breakdown spectroscopy using wood slice substrates Zhi-jiang Chen, Hong-kun Li, Ming Liu, Run-hua Li Liquids Hamamatsu CR114 Photomultiplier Tianjin Tuopu WDS-5
Laser: Nd:YAG
1064.0000nm
100.0000mJ
5.000Hz
Gate Delay: 10.000us
Gate Width: 5.000us
A thin slice of wood (1mm thick) absorbs the liquid to be analysed (2 mins), is dried for 1 min and then can be LIBS analysed. 4.5% RSD achieved using pulse averaging. Technique has the disadvantage that elements in the wood at appreciable concentrations (Ca, Sr, Na) cannot be analysed.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
Cr 0.0340 (Calibration Curve) 425.4300 N/A Univariate -10.0000--10.0000 Liquid sample absorbed into wood slice for analysis
Cu 0.0290 (Calibration Curve) 324.7500 N/A Univariate -10.0000--10.0000 Liquid sample absorbed into wood slice for analysis
Cd 0.5900 (Calibration Curve) 508.5800 N/A Univariate -10.0000--10.0000 Liquid sample absorbed into wood slice for analysis
Pb 0.0740 (Calibration Curve) 405.7800 N/A Univariate -10.0000--10.0000 Liquid sample absorbed into wood slice for analysis
Mn 0.0360 (Calibration Curve) 255.1200 N/A Univariate -10.0000--10.0000 Liquid sample absorbed into wood slice for analysis