Quantitative analysis of trace metal ions in ice using laser-induced breakdown spectroscopy J.O. Caceres, J. Tornero Lopez, A. Gonzalez Urena, H.H. Telle Liquids EG&G 1461/88 Not Known
Laser: Nd:YAG
532.0000nm
200.0000mJ
10.000Hz
Gate Delay: 4.000us
Gate Width: 5.000us
Some experiments also carried out using a Q-Switched CO2 laser. Liquid sample analysis was simplified by freezing the samples and analysing the solid.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
Al 1.0000 (Stated) -10.0000 N/A Univariate Calibration 200.0000-5000.0000 Peak intensity ratioed to background signal (no further details supplied)
Na 2.0000 (Stated) 588.9950 N/A Univariate Calibration 200.0000-5000.0000 Peak ratioed to signal background


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes