ENHANCED SENSITIVITY OF DIRECT BERYLLIUM DETERMINATION IN SOIL BY LASER-INDUCED BREAKDOWN SPECTROMETRY T.A. Labutin, N.B. Zorov, S.M. Zaytsev, M.O. Kozhnov, A.M. Popov Soil ICCD Jobin-Yvon HRP
Laser: Nd:YAG
355.0000nm
60.0000mJ
NoneHz
Gate Delay: 0.750us
Gate Width: 1.000us
Double Pulse LIBS was used to analyse beryllium in samples made by compressing soil into pellets. The 313.04nm line was preferred due to interferences at the other strong lines (234.86nm, interference with Fe; 332.11nm interference with Ti). Detection limit for 355nm was a factor of 3 better than with 532nm laser.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
Be 0.0700 (Calibration Curve Slope) 313.0400 N/A Univariate -10.0000--10.0000 DP LIBS. 355nm source laser.
Be 0.2000 (Calibration Curve Slope) 313.0400 N/A Univariate -10.0000--10.0000 Double pulse LIBS. Detection limit using 532nm