LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
ENHANCED SENSITIVITY OF DIRECT BERYLLIUM DETERMINATION IN SOIL BY LASER-INDUCED BREAKDOWN SPECTROMETRY | T.A. Labutin, N.B. Zorov, S.M. Zaytsev, M.O. Kozhnov, A.M. Popov | Soil | ICCD | Jobin-Yvon HRP | |
Laser: | Nd:YAG 355.0000nm 60.0000mJ NoneHz |
||||
Gate Delay: | 0.750us | ||||
Gate Width: | 1.000us | ||||
Double Pulse LIBS was used to analyse beryllium in samples made by compressing soil into pellets. The 313.04nm line was preferred due to interferences at the other strong lines (234.86nm, interference with Fe; 332.11nm interference with Ti). Detection limit for 355nm was a factor of 3 better than with 532nm laser. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
Be | 0.0700 (Calibration Curve Slope) | 313.0400 | N/A | Univariate | -10.0000--10.0000 | DP LIBS. 355nm source laser. |
Be | 0.2000 (Calibration Curve Slope) | 313.0400 | N/A | Univariate | -10.0000--10.0000 | Double pulse LIBS. Detection limit using 532nm |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|