LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Field Deployable Method for Gold Detection Using Gold Pre-Concentration on Functionalized Surfaces | Heike Ebendorff-Heidepriem, Krasimir Vasilev, Caroline Tiddy, Steven Tassios, Akash Bachhuka, Agnieszka Zuber | Geological | Spectrolaser 4000 | Spectrolaser 4000 | Spectrolaser |
Laser: | Nd:YAG 532.0000nm 110.0000mJ NoneHz |
||||
Gate Delay: | 0.800us | ||||
Gate Width: | 1000.000us | ||||
The authors demonstrate a method for pre-concentrating gold onto a planar substrate, which is then analysed using LIBS. This has a applications for field analysis, without the use of dangerous chemicals. Tests with geological sample showed good leaching of gold from the sample (47% or greater). LIBS analysis of rock powder samples failed to detect Au (even at 47ppm) due to the low chance of scanning the rock particles bearing gold. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
Au | 0.0800 (Minimum Detectable Concentration) | 267.6000 | 274.8 | None | -10.0000--10.0000 | Using solutions only containing gold |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|