LIBS Info: Element Analysis

Title Authors Material Detector Spectrometer Software
Direct determination of Ca, K, Mg, Na, P, S, Fe and Zn in bivalve mollusks by wavelength dispersive X-ray fluorescence (WDXRF) and laser-induced breakdown spectroscopy (LIBS) Diego Victor de Babos, Edenir Rodrigues Pereira-Filho, Fabio Alan Carqueija Amorim, Vinicius Camara Costa Biological Applied Spectra J200 Applied Spectra J200 Microsoft Excel
Laser: Nd:YAG
1064.0000nm
40.0000mJ
5.000Hz
Gate Delay: 1.000us
Gate Width: 1050.000us
The authors compare the performance of Wavelength Dispersive XRF with LIBS for the analysis of elements in Molluscs. LIBS analysis used pellets made from freeze dried mollusc samples. Extensive work was undertaken to determine the best conditions for sample analysis with the Standard Error of Calibration being the optimised parameter.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
Na 20.0000 (Calibration Curve Slope) 588.9900 N/A Univariate -10.0000--10.0000 Analysis of dried mollusc pellets after extensive parameter optimisation
Mg 10.0000 (Calibration Curve Slope) 285.2100 N/A Univariate -10.0000--10.0000 Analysis of dried mollusc pellets after extensive parameter optimisation
K 5.0000 (Calibration Curve Slope) 766.4900 N/A Univariate -10.0000--10.0000 Analysis of pellets of dried mollusc
Ca 7.0000 (Calibration Curve Slope) 527.0500 N/A Univariate -10.0000--10.0000 Result of optimising experimental conditions


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes
Na 0.9300 % 588.9900 Univariate Calibration Analysis of dried mollusc pellets after extensive parameter optimisation. Average of 13 samples. By comparison, WDXRF RMSEP=0.18%
Mg 0.0680 % 285.2100 Univariate Calibration Analysis of dried mollusc pellets after extensive parameter optimisation. Average of 28 samples. By comparison, WDXRF RMSEP=0.043%
K 0.0680 % 766.4900 Univariate Calibration Analysis of dried mollusc pellets after extensive parameter optimisation. Average of 28 samples. By comparison, WDXRF RMSEP=0.063%
Ca 0.0690 % 527.0500 Univariate Calibration Average over 28 samples. By comparison, XRF (Wavelength Dispersive) had RMSEP: 0.043%