LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Analysis of Limestone using the TSI LIBS Desktop Analyzer | N/A | Limestone | Spectrolaser 1000 | Spectrolaser 1000 | Spectrolaser |
Laser: | Nd:YAG 1064.0000nm NonemJ NoneHz |
||||
Gate Delay: | Noneus | ||||
Gate Width: | Noneus | ||||
Results of analysis of pressed powders. App Note suggests <5% RSD for most analyses. Detectable elements include (but not confined to) Al, Ba, C, Ca, Cr, Fe, H, Mg, Mn, Na, O, Si, Sr and Ti. Detection Limit also provided for Na (10ppm) and Ti (20ppm) but no wavelength or calibration curve supplied. Analysis results for S using PLS shown, but no further explanation. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
Mg | 10.0000 (Stated) | 293.0000 | N/A | Calibration Curve | None-None | No working/explanation as to how DT was calculated/derived. No Calibration curve supplied. |
Al | 30.0000 (Stated) | 309.2700 | N/A | Calibration Curve | None-None | No working/explanation as to how DT was calculated/derived. No Calibration curve supplied. |
Si | 30.0000 (Stated) | 263.1000 | 288.157 | Calibration Curve | 0.0000-140000.0000 | No working/explanation as to how DT was calculated/derived. Calibration curve shown for 288.157nm, although results given for 263.1nm. |
K | 20.0000 (Stated) | 766.4910 | N/A | Calibration Curve | 0.0000-14000.0000 | No working/explanation as to how DT was calculated/derived. |
Ca | 10.0000 (Stated) | 643.9070 | N/A | Calibration Curve | 700000.0000-800000.0000 | No working/explanation as to how DT was calculated/derived. Calibration curve supplied, but does not show values close to 0% |
Fe | 30.0000 (Stated) | 259.9390 | N/A | Calibration Curve | 0.0000-18000.0000 | No working/explanation as to how DT was calculated/derived. |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|