LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Hydrogen Leak Detection Using Laser-Induced BreakdownSpectroscopy | D.W. Hahn, V. Hohreiter, A.J. Ball | Gases | ICCD | Czerny-Turner (Unspecified) | |
Laser: | Nd:YAG 1064.0000nm 55.0000mJ 2.500Hz |
||||
Gate Delay: | 4.000us | ||||
Gate Width: | 2.000us | ||||
The authors explore 2 different methods for the detection of low concentrations of H gas - a LIBS spark formed on a stainless steel surface in the test atmosphere and a LIBS spark formed directly in the test gas atmosphere |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
H | 23.0000 (Calibration Curve Slope) | 656.0000 | N/A | Univariate | 65.0000-820.0000 | 100mJ Pulse used to analyse test gas directly |
H | 15.0000 (Calibration Curve Slope) | 656.0000 | N/A | Univariate | 65.0000-820.0000 | 50mJ pulse on a stainless steel substrate in the test atmosphere |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|