LIBS Info: Element Analysis

Tap
Save
Hover
Title Authors Material Detector Spectrometer Software
Resonance/non-resonance doublet-based selfabsorption- free LIBS for quantitative analysis with a wide measurement range S.T. Jia, L.T. Xiao, W.B. Yin, L. Dong, W.G. Ma, Yun Zhao, L. Zhang, J.J. Hou Brass Princeton Instruments ICCD Max Princeton Instruments Grating Spectrometer
Laser: Nd:YAG
1064.0000nm
50.0000mJ
20.000Hz
Gate Delay: 0.200us
Gate Width: 0.100us
Samples were Copper Oxide/Potassium Bromide pressed pellets in the range of 0.01 - 60%. The Authors use a combination of Resonance Lines [324.7nm] and non-resonance lines [521nm] to produce a combined calibration curve. The curve has different linear segments depending on the concentration of the sample under evaluation.