LIBS Info: Element Analysis

Title Authors Material Detector Spectrometer Software
Feasibility of using laser induced breakdown spectroscopy for quantitative measurement of calcium, magnesium, potassium and sodium in meat Cassiana Seimi Nomura, Angerson Nogueira Nascimento, Daniel Menezes Silvestre, Flavio O. Leme Organic J 200 Tandem J 200 Tandem Aurora
Laser: Nd:YAG
266.0000nm
20.0000mJ
10.000Hz
Gate Delay: 0.200us
Gate Width: 1050.000us
A rare application where ground beef and chicken samples are analysed using LIBS. Samples were freeze dried, then milled, then pelletised. Elements for analysis were introduced via the cellulose binder. Laser spot size, and delay were optimised for analysis.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
Na 11.8000 (Calibration Curve Slope) 588.9950 589.592nm Univariate 280.0000-2300.0000 Detection limits in the range 11.8-13.7 ppm were obtained depending on material (Beef or Chicken) and wavelength used.
K 25.9000 (Calibration Curve Slope) 766.4890 769.896nm Univariate 1400.0000-16000.0000 Detection limits in the range 25.9-38.6 ppm were obtained depending on material (Beef or Chicken) and wavelength used.
Mg 7.2000 (Calibration Curve Slope) 285.2130 279.553nm, 280.27nm Univariate 220.0000-1500.0000 Detection limits in the range 7.2-10.5 ppm were obtained depending on material (Beef or Chicken) and wavelength used.
Ca 7.0000 (Calibration Curve Slope) 422.0000 393.366nm, 396.847nm Univariate 75.0000-480.0000 Detection limits in the range 7-10.7 ppm were obtained depending on material (Beef or Chicken) and wavelength used.


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes
Ca 16.6883 ppm 422.6730 Univariate Analysis of pelletised, ground beef/chicken. Analysis of pelletised, ground beef/chicken.
K 588.3026 ppm 766.4890 Univariate Analysis of pelletised, ground beef/chicken. Analysis of pelletised, ground beef/chicken.
Mg 22.8254 ppm 285.2130 Univariate Analysis of pelletised, ground beef/chicken. Analysis of pelletised, ground beef/chicken.
Na 62.1168 ppm 588.9950 Univariate Analysis of pelletised, ground beef/chicken. Analysis of pelletised, ground beef/chicken.