Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Carbon Nanotube Analysis using the TSI LIBS Desktop Analyzer | N/A | Carbon Nanotubes | ChemReveal | ChemReveal | |
Laser: | Nd:YAG 1064.0000nm 30.0000mJ NoneHz |
||||
Gate Delay: | 1.000us | ||||
Gate Width: | Noneus | ||||
Metal Impurities (Ni, Co, Fe) in Carbon Nanotubes (CNT) are analyzed. Pellets are constructed using different mixtures of 2 different CNTs. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
Ni | 1000.0000 (Lowest Calib Conc) | 227.0000 | N/A | Calibration Curve | 1000.0000-18000.0000 | Note: LOD listed is the lowest calibration point. Actual LOD may be lower. |
Co | 2000.0000 (Lowest Calib Conc) | 238.0000 | N/A | Calibration Curve | 2000.0000-30000.0000 | Note: LOD listed is the lowest calibration point. Actual LOD may be lower. |
Fe | 200.0000 (Lowest Calib Conc) | 239.0000 | N/A | Calibration Curve | 200.0000-6000.0000 | Note: LOD listed is the lowest calibration point. Actual LOD may be lower. |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|