Quantitative LIBS analysis of vanadium in samples of hexagonal mesoporous silica catalysts | M. Pouzar, T. Kratochvil, Libor Capek, Lucie Smolakova, T. Cernohorsky, Anna Krejcova, Ludek Hromadko | Metals | SOLAR LEA-S500 | SOLAR LEA-S500 | |
Laser: | Nd:YAG 1064.0000nm 95.0000mJ 20.000Hz |
||||
Gate Delay: | Noneus | ||||
Gate Width: | Noneus | ||||
This result utilises Double Pulse LIBS with a inter-pulse delay of 7us. RMSE results reflect using a calibration based on Vanadium - Hexagonal Mesoporous Silica [V-HMS] (impregnated) to analyse both V-HMS (impregnated) and V-HMS (synthesised samples). However, V-HMS synthesised samples have Vanadium in a different form, which is not similar enough for a good result. |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
V | 1300.0000 (Calibration Curve Slope) | 310.2300 | 311.071nm, 311.838nm | Univariate | -10.0000--10.0000 | Sample was spread on adhesive tape for analysis |
Element | RMSE (ppm) | Wavelength (nm) | Calibration Method | Notes |
---|---|---|---|---|
V | 14366.7440 ppm | 310.2300 | Univariate Calibration | Reference Conc from ED-XRF |