Title Authors Material Detector Spectrometer Software
Quantitative LIBS analysis of vanadium in samples of hexagonal mesoporous silica catalysts M. Pouzar, T. Kratochvil, Libor Capek, Lucie Smolakova, T. Cernohorsky, Anna Krejcova, Ludek Hromadko Metals SOLAR LEA-S500 SOLAR LEA-S500
Laser: Nd:YAG
1064.0000nm
95.0000mJ
20.000Hz
Gate Delay: Noneus
Gate Width: Noneus
This result utilises Double Pulse LIBS with a inter-pulse delay of 7us. RMSE results reflect using a calibration based on Vanadium - Hexagonal Mesoporous Silica [V-HMS] (impregnated) to analyse both V-HMS (impregnated) and V-HMS (synthesised samples). However, V-HMS synthesised samples have Vanadium in a different form, which is not similar enough for a good result.
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
V 1300.0000 (Calibration Curve Slope) 310.2300 311.071nm, 311.838nm Univariate -10.0000--10.0000 Sample was spread on adhesive tape for analysis


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes
V 14366.7440 ppm 310.2300 Univariate Calibration Reference Conc from ED-XRF