LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Calibration free laser-induced breakdown spectroscopy of oxide materials | B. Praher, V. Palleschi, R. Viskup, J. Heitz, J.D. Pedarnig | Oxides | ARYELLE | ARYELLE | MATLAB |
Laser: | Nd:YAG 1064.0000nm 90.0000mJ 5.000Hz |
||||
Gate Delay: | 0.800us | ||||
Gate Width: | 0.200us | ||||
CF LIBS analysis was performed on pellets consisting of different concentrations of 3 different oxide materials - Fe203, MgO and CaO. The authors systematically work on the aspects spectra/experiment that deviate from the 'ideal' - background (by collecting at a delay that has effectively no background), noise (by averaging 500 spectra), spectral response of detector (by measuring independently and correcting), deconvolution of the instrument function (by measurements using Hg lamp) and self absorption (by choosing lines with lower levels sufficiently (according to a detailed criteria)above the ground state) |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|