Title Authors Material Detector Spectrometer Software
Analyzing Glass with Laser Induced Breakdown Spectroscopy N/A Glass ChemReveal ChemReveal
Laser: Nd:YAG
266.0000nm
50.0000mJ
NoneHz
Gate Delay: Noneus
Gate Width: Noneus
Application note demonstrating the use of LIBS for the analysis of components of glass. The 266nm laser is necessary to get reliable breakdown of the material (material is too transparent at visible and infrared wavelengths). Analysis performed in Argon atmosphere
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
B 3000.0000 (Minimum Concentration Detected) 249.7740 N/A Univariate 3000.0000-110000.0000 Argon atmosphere. Peak ratioed to Si (maybe 250.69nm)


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes
Na 2458.6582 ppm -10.0000 Univariate Calibration
Mg 618.4658 ppm -10.0000 Univariate Calibration
Li 9.0000 ppm -10.0000 Univariate Calibration
Ca 785.8117 ppm -10.0000 Univariate Calibration