LIBS Info: Element Analysis

Title Authors Material Detector Spectrometer Software
Laser-induced breakdown spectroscopy for simultaneous determination of Sm, Eu and Gd in aqueous solution Suresh K. Aggarwal, D. Alamelu, Arnab Sarkar Lanthanides Spectrolaser 1000 Spectrolaser 1000 Spectrolaser
Laser: Nd:YAG
1064.0000nm
60.0000mJ
10.000Hz
Gate Delay: 3.200us
Gate Width: 1000.000us
Before analysis 40micolitres of liquid sample (to which the LODs relate) where drop-wise transferred to Whatman 42 filter paper, then evaporated to dryness. [thus, providing a concentrating step]. Both LODs and RMSE data determined. Peaks show some Self Absorption, but lower part of each calibration is linear
Element Detection Limit (ppm) Wavelength (nm) Other Wavelengths (nm) Calibration Method Calibration Range (ppm) Notes
Sm 1.3000 (Calibration Curve Slope) -10.0000 471.706 Univariate 50.0000-2000.0000 Concentration in original solution, analysis performed on sample transferred to filter paper and dried.
Eu 1.9000 (Calibration Curve Slope) 462.7720 N/A Univariate 50.0000-2000.0000 Concentration in original solution, analysis performed on sample transferred to filter paper and dried.
Gd 2.3000 (Calibration Curve Slope) 336.2230 N/A Univariate 50.0000-2000.0000 Concentration in original solution, analysis performed on sample transferred to filter paper and dried.


Element RMSE (ppm) Wavelength (nm) Calibration Method Notes
Sm 21.7945 ppm 471.7060 Univariate Calibration
Eu 17.3205 ppm 462.7720 Univariate Calibration
Gd 15.8114 ppm 336.2230 Univariate Calibration