LIBS Info: Element Analysis
Title | Authors | Material | Detector | Spectrometer | Software |
---|---|---|---|---|---|
Laser-induced breakdown spectroscopy for simultaneous determination of Sm, Eu and Gd in aqueous solution | Suresh K. Aggarwal, D. Alamelu, Arnab Sarkar | Lanthanides | Spectrolaser 1000 | Spectrolaser 1000 | Spectrolaser |
Laser: | Nd:YAG 1064.0000nm 60.0000mJ 10.000Hz |
||||
Gate Delay: | 3.200us | ||||
Gate Width: | 1000.000us | ||||
Before analysis 40micolitres of liquid sample (to which the LODs relate) where drop-wise transferred to Whatman 42 filter paper, then evaporated to dryness. [thus, providing a concentrating step]. Both LODs and RMSE data determined. Peaks show some Self Absorption, but lower part of each calibration is linear |
Element | Detection Limit (ppm) | Wavelength (nm) | Other Wavelengths (nm) | Calibration Method | Calibration Range (ppm) | Notes |
---|---|---|---|---|---|---|
Sm | 1.3000 (Calibration Curve Slope) | -10.0000 | 471.706 | Univariate | 50.0000-2000.0000 | Concentration in original solution, analysis performed on sample transferred to filter paper and dried. |
Eu | 1.9000 (Calibration Curve Slope) | 462.7720 | N/A | Univariate | 50.0000-2000.0000 | Concentration in original solution, analysis performed on sample transferred to filter paper and dried. |
Gd | 2.3000 (Calibration Curve Slope) | 336.2230 | N/A | Univariate | 50.0000-2000.0000 | Concentration in original solution, analysis performed on sample transferred to filter paper and dried. |